Suitable for safe processing of microbiological cultures. The INCUCELL V line (with a fan) ensures quicker and more precise temperature distribution with small deviations. These units are to be used especially in biological and microbiological laboratories. They can also be used for quality tests in the pharmaceutical industry and cosmetics, for testing in the veterinary medicine, and in the food industry.
- Inner volume of the chamber: 22, 55, 111, 222, 404, and 707 liters
- Working temperature: From 10°C above ambient temperature up to 100°C
- Inner glass door
- Inner chamber: Stainless steel, DIN 1.4301 (AISI 304)
- Natural, precise – laboratory incubator
- 111-liter chamber volume
- Working temperature of from +10°C above ambient up to 100°C
- Patented system of air flow in the chamber ensures quicker and more accurate distribution of the preset temperature in the chamber, which enables ideal heating of the samples
- High operation comfort, precise temperature regulation and short times of temperature recovery in the chamber after the door opening; the standard equipment includes inner glass door
- Model with a fan ensures quicker and more accurate distribution of temperature with small deviations; suitable for demanding and accurate tests and samples cultivation
- Model with a Fuzzy logic microprocessor control system; control elements situated on a foil keyboard, process information indicated on a LED display
- Wide range of optional equipment and accessories
- Output for a printer or PC connection
- Validation possibility (IQ , OQ)
Technical data
Interior of stainless steel material No. 1.4301
- volume111 l
- depth370 mm
- width540 mm
- height530 mm
Tray
- standard equipment2 pcs
- storage area (w × d)520×335 mm
- racks7 max. number
- distance between guide rails70 mm
Maximal weight of the load *)
- inside the oven50 kg/oven
- per tray20 kg/tray
Door
- 1 pcs
External dimensions (including door and handle)
- depth640 mm
- diameter of the air branch outer/inner52/49 mm
- width760 mm
- height, incl. Feet (N) and Rolls (K)860 (N)
Package dimensions (three layers carton)
- depth730 mm
- width830 mm
- depth1060 mm
Weight
- brut87 kg
- nett75 kg
Working temperature (beginning of the regulation)
- from 10 °C above ambient temp. to:100 °C
Temperature accuracy according to DIN 12 880 T2 at working temperature with closed air flap and door
- time variation cca (±)<=0,2 °C
- space deviation cca (±)<=0,3 °C
Time required to reach 37 °C with closed air flap and voltage 230 V
- 51 min
Heat emission at 37 °C
- 95 W
Air exchange speed at 37 °C
- 49 per hour
Electricity - mains 50/60 Hz
- max. power0,7 kW
- nominal voltage230 V
- current3 A
- power by stand by5 W
Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.