İnkübatörler
MMM INCUCELL 222 - ECO line
Suitable for safe treatment of microbiological cultures. The INCUCELL line produces no noise and provides a very soft air flow within the chamber. These devices can be used especially in biological and microbiological laboratories, quality tests in pharmaceutical and cosmetic industry, and for testing in veterinary medicine and food industry.
- Internal volume: 22, 55, 111, 222, 404, and 707 liters
- Working temperature range: from 5°C above ambient temperature up to 100°C
- Inner glass door
- Inner chamber: Stainless steel DIN 1.4301 (AISI 304)
- Natural, precise – laboratory incubator
- 222-liter chamber volume
- Working temperature of from +5°C above ambient up to 100°C
- Noiseless operation and fine air flow in the chamber
- High operation comfort, precise temperature regulation and short times of temperature recovery in the chamber after the door opening
- High-quality and safe drying, heating and tempering of common laboratory materials in laboratories and industry; suitable for safe processing of microbiological cultures
- Model with a Fuzzy logic microprocessor control system; control elements situated on a foil keyboard, process information indicated on a LED display
- Wide range of optional equipment and accessories
- Output for a printer or PC connection
- Validation possibility (IQ , OQ)
Technical data
Interior of stainless steel material No. 1.4301
- volume222 l
- depth520 mm
- width540 mm
- height760 mm
Tray
- standard equipment2 pcs
- storage area (w × d)520×485 mm
- racks10 max. number
- distance between guide rails70 mm
Maximal weight of the load*)
- inside the oven70 kg/oven
- per tray30 kg/tray
Door
- 1 pcs
External dimensions (including door and handle)
- depth790 mm
- diameter of the air branch outer / inner52/49 mm
- width760 mm
- height, incl. Feet (N) and Rolls (K)1095 (N)
Package dimensions (three layers carton)
- depth860 mm
- width860 mm
- depth1260 mm
Weight
- brut116 kg
- nett100 kg
Working temperature (beginning of the regulation)
- from 5 °C above ambient temp. to:100 °C
Temperature accuracy according to DIN 12 880 T2 at working temperature with closed air flap and door
- time variation cca (±)<=0,2 °C
- space deviation cca (±)< 1 °C
Time required to reach 37 °C with closed air flap and voltage 230 V
- 79 min
Heat emission at 37 °C
- 45 W
Air exchange speed at 37 °C
- 5 per hour
Electricity - mains 50/60 Hz
- max. power0,5 kW
- nominal voltage230 V
- current2 A
- power by stand by5 W
*) Approx. 50 % of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber. Note: All technical data are related to 22 °C ambient temperature and +-10 % voltage swing (if not specified).