Etüvler - Laboratuvar Fırınları
MMM Ecocell 707 - ECO line
Convincingly cost saving laboratory dryer 707-liter chamber volume Working temperature of from +5°C over ambient temperature up to 250/300°C Natural fine air flow in the chamber, noiseless operation High-quality and safe drying, heating and tempering of common laboratory materials in...
Technical data
Interior of stainless steel material No. 1.4301
- volume707 l
- depth520 mm
- width940 mm
- height1415 mm
Tray
- standard equipment2 pcs
- storage area (w × d)920×485 mm
- racks19 max. no
- distance between guide rails70 mm
Maximal weight of the load *)
- inside the oven130 kg/oven
- per tray50 kg/tray
Door
- 2 pcs
External dimensions (including door and handle)
- depth790 mm
- diameter of the air branch outer/inner52/49 mm
- width1160 mm
- height, incl. Feet (N) and Rolls (K)1910 (K)
Package dimensions (three layers carton)
- depth860 mm
- width1230 mm
- depth2085 mm
Weight
- brut240 kg
- nett215 kg
Working temperature (beginning of the regulation)
- from 5 °C above ambient temp. to:250 °C
Temperature accuracy according to DIN 12 880 T2, at working temperature with closed air flap and door
- time variation cca (±)<= 1 °C
- space deviation cca (±) % of the reached temperature3,5 %
Time required to reach 250 °C with closed air flap and voltage 230 V
- 95 min
Heat emission at 250 °C
- 2550 W
Air exchange speed at 150 °C
- 3 per hour
Electricity - mains 50/60 Hz
- max. power input5,4 kW
- nominal voltage400/3N V
- current7,8 A
- input in stand by mode5 W
Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.
Toplam Görüntülenme 2178