Etüvler - Laboratuvar Fırınları
MMM Durocell 22 - ECO line
Highly durable laboratory oven 22-liter chamber volume Working temperature of from +5°C over ambient temperature up to 125°C Natural fine air flow in the chamber, noiseless operation High-quality and safe drying, heating and tempering of laboratory materials in laboratories and industry;...
Technical data
Interior of stainless steel material No. 1.4301
- volume22 l
- depth350 mm
- width240 mm
- height300 mm
Trays
- standard equipment2 pcs
- storage area (w × d)185×265 mm
- racks4 max. no
- distance between guide rails60 mm
Maximal weight of the load*)
- inside the oven25 kg/oven
- per tray10 kg/tray
Door
- 1 pcs
External dimensions (including door and handle)
- depth560 mm
- diameter of the air branch outer/inner52/49 mm
- width500 mm
- height (inclusive foots - N)610 (N) mm
Package dimensions (three layers carton)
- depth720 mm
- width500 mm
- depth810 mm
Weight
- brut36 kg
- nett31 kg
Working temperature (beginning of the regulation)
- from 5 °C above ambient temp. to:125 °C
Temperature accuracy according to DIN 12 880 T2, at working temperature with closed air flap and door
- time variation cca (±)1 °C
- space deviation cca (±) % of the reached temperature2,7 %
Time required to reach 100 °C with closed air flap and voltage 230 V
- 34 min
Heat emission at 100 °C
- 140 W
Air exchange speed at 100 °C
- 6 per hour
Electricity – mains 50/60 Hz
- max. power input0,9 kW
- nominal voltage230 V
- current4 A
- input in stand by mode5 W
Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.
Toplam Görüntülenme 2342