Etüvler - Laboratuvar Fırınları
MMM Ecocell 55 - ECO line
Convincingly cost saving laboratory dryer 55-liter chamber volume Working temperature of from +5°C over ambient temperature up to 250/300°C Natural fine air flow in the chamber, noiseless operation High-quality and safe drying, heating and tempering of common laboratory materials in...
Technical data
Interior of stainless steel material No. 1.4301
- volume55 l
- depth370 mm
- width400 mm
- height350 mm
Tray
- standard equipment2 pcs
- storage area (w × d)380×335 mm
- racks4 max. no
- distance between guide rails70 mm
Maximal weight of the load *)
- inside the oven50 kg/oven
- per tray20 kg/tray
Door
- 1 pcs
External dimensions (including door and handle)
- depth640 mm
- diameter of the air branch outer/inner52/49 mm
- width620 mm
- height, incl. Feet (N) and Rolls (K)680 (N)
Package dimensions (three layers carton)
- depth730 mm
- width700 mm
- depth875 mm
Weight
- brut66 kg
- nett55 kg
Working temperature (beginning of the regulation)
- from 5 °C above ambient temp. to:250 °C
Temperature accuracy according to DIN 12 880 T2, at working temperature with closed air flap and door
- time variation cca (±)0,3 °C
- space deviation cca (±) % of the reached temperature2 %
Time required to reach 250 °C with closed air flap and voltage 230 V
- 59 min
Heat emission at 250 °C
- 590 W
Air exchange speed at 150 °C
- 8 per hour
Electricity - mains 50/60 Hz
- max. power input1,2 kW
- nominal voltage230 V
- current5,2 A
- input in stand by mode5 W
Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.